inforMED
MalfunctionNHW

PATCH

Received Jan 7, 2011 · Event occurred Dec 7, 2010

Report 2124215-2010-23681 · MDR key 1952756

Device

Generic name

Implantable Lead

Model number

0067

Product problems

  • High impedance

Patient

Not reported

  • Other (for use when an appropriate patient code cannot be identified)

Narrative

Additional Manufacturer Narrative

THE LEADS REMAIN IMPLANTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED AND AN AMENDED REPORT SUBMITTED.

Description of Event or Problem

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THREE MONTHS AGO, THE SHOCK IMPEDANCE VALUES WERE HIGHER THAN 125 OHMS. ON (B)(6) 2010 THE VALUES WERE BETWEEN 60 TO 70 OHMS. THE PATIENT HAS TWO EPICARDIAL DEFIBRILLATION LEADS IMPLANTED (MODELS 0067 AND 0068, SERIAL NUMBERS UNKNOWN) AND A COMPETITOR PACE/SENSE LEAD. A SHOCK TEST WAS NOT PERFORMED DUE TO THE PATIENTS POOR CONDITION. THE LEADS WILL REMAIN IMPLANTED AND MONITORED CLOSELY EVERY THREE MONTHS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.