inforMED
MalfunctionMPO

INFANT HEEL WARMER SQUEEZE 4X4IN

Received Jun 11, 2024 · Event occurred May 15, 2024

Report 1423537-2024-00029 · MDR key 19515762

Device

Generic name

Infant Heel Warmer (chemical Heat Pack)

Manufacturer

Mo016 Mo-moberly

Model number

11460-010T

Catalog number

11460-010T

Lot number

V2L195

Product problems

  • Burst Container or Vessel
  • Burst Container or Vessel

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

EXPLODED IN THEIR HANDS WHILE THEY WERE WARMING INFANTS HEEL.

Additional Manufacturer Narrative

ONE UN-USED SAMPLE WAS RETURNED FOR EVALUATION AND ACTIVATED IN THE LAB WITHOUT ISSUE. AS NO ISSUES WERE FOUND DURING INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE DEVICE HISTORY RECORD REVIEW WAS COMPLETED ON THE REPORTED LOT NUMBER- V2E092, AND THE LOT WAS MANUFACTURED AND RELEASED IN COMPLIANCE WITH ALL REQUIREMENTS. CARDINAL HEALTH WILL CONTINUE TO MONITOR COMPLAINT TRENDS FOR THIS REPORTED ISSUE.