GEM MICROVASCULAR ANASTOMOTIC COUPLER
Received Jun 10, 2024 · Event occurred May 14, 2024
Report 1416980-2024-02822 · MDR key 19504510
Device
Generic name
Device, Anastomotic, Microvascular
Manufacturer
Baxter Healthcare CorporationModel number
GEM2753 ICatalog number
511100250010
Lot number
SP23115-1940086
Product problems
- Disconnection
- Disconnection
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE RINGS FELL OUT OF THE WING JAW ASSEMBLY OF A 2.5MM COUPLER. THIS WAS OBSERVED AFTER THE PRODUCT PACKAGE WAS OPENED, PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Additional Manufacturer Narrative
D4: THE LOT # AND EXPIRATION DATE WERE NOT INCLUDED IN THE UDI # AS THE LOT WAS NOT RECOGNIZED IN THE SYSTEM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: H3, H4, H6, H11. H11: TWO DEVICES WERE RECEIVED FOR EVALUATION. THE FIRST JAW ASSEMBLY WAS REVIEWED FOR THE REPORTED PROBLEM CODE OF ¿RING DISLODGEMENT¿. THE RIGHT RING OF THE 2.5MM COUPLER JAW ASSEMBLY WAS NO LONGER SEATED IN THE JAW ASSEMBLY WHICH WOULD INDICATE THAT THE RING HAD DISLODGED FROM THE JAW ASSEMBLY. THERE WERE NO SIGNS OF USE ON THE REMAINING COUPLER RING (DRIED BLOOD/TISSUE) WHICH WOULD INDICATE THAT THE ALLEGED EVENT OCCURRED PRIOR TO USE IN A SURGICAL PROCESS. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A DEVICE HISTORY REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.