inforMED
MalfunctionHSH

NEXGEN ARTICULAR SURFACE INSERTER

Received Dec 9, 2010 · Event occurred Nov 8, 2010

Report 1822565-2010-01319 · MDR key 1950203

Device

Generic name

Knee Instrument

Manufacturer

Zimmer Inc

Catalog number

00597702000

Lot number

60582624

Product problems

  • Fracture

Patient

Not reported

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

IT IS REPORTED THAT THE SURFACE INSERTER HAS FRACTURED.

Additional Manufacturer Narrative

THE FRACTURE LIKELY OCCURRED DUE TO THE INCORRECT MATERIAL BEING UTILIZED FOR THIS PRODUCTION LOT. THIS LOT HAS BEEN RECALLED, SEE 1822565-05/19/08-001R FOR ADDITIONAL DETAILS. DIMENSIONAL READING IS CONFORMING TO PRINT SPECIFICATION. MATERIAL HARDNESS IS NONCONFORMING. AS RETURNED, THE TIP OF THE ARTICULAR SURFACE INSERTER HAS FRACTURED AND WAS NOT RETURNED FOR REVIEW. THE DEVICE HAS A POTENTIAL FIELD AGE OF APPROXIMATELY 4 YEARS.

Remedial action

  • Recall