ADVIA CHEMISTRY
Received Jun 6, 2024 · Event occurred Apr 25, 2024
Report 2432235-2024-00128 · MDR key 19475878
Device
Generic name
Advia Chemistry Urinary/cerebrospinal Fluid Protein (ucfp)
Manufacturer
Siemens Healthcare Diagnostics Inc.Catalog number
SMN 11319151
Lot number
140
Product problems
- Low Test Results
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED AN ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULT ON AN ADVIA® 2400 CLINICAL CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN AND WAS NOT QUESTIONED. THE SAME SAMPLE WAS REPROCESSED ON THE SAME ADVIA® 2400 CLINICAL CHEMISTRY SYSTEM. THE REPROCESSED RESULT OBTAINED WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT.
Additional Manufacturer Narrative
AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING AN ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULT OBTAINED ON AN ADVIA® 2400 CLINICAL CHEMISTRY SYSTEM. SIEMENS IS INVESTIGATING THE EVENT.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION (11-JUL-2024): SIEMENS HEALTHCARE DIAGNOSTICS CONCLUDED THE INVESTIGATION OF THE EVENT. SIEMENS CONFIRMED THE POTENTIAL FOR BIASED QUALITY CONTROL (QC) AND PATIENT RESULTS WHEN USING ADVIA CHEMISTRY URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) LOT 140 ON THE ADVIA® 2400 CHEMISTRY SYSTEM. SIEMENS¿ INTERNAL INVESTIGATION CONFIRMED WHEN USING LOT 140, QC RECOVERED OUTSIDE OF THE ALLOWABLE CONTROL LIMITS FOR URINE CHEMISTRY CONTROL LEVELS. UNITED STATES (US) CUSTOMERS WERE NOTIFIED THROUGH AN URGENT MEDICAL DEVICE CORRECTION (UMDC) ACHC24-04.A.US.CHC, TITLED "ADVIA CHEMISTRY URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) LOT 140 QUALITY CONTROL (QC) OUT OF RANGE AND BIASED PATIENT RESULTS" AND OUTSIDE THE UNITED STATES (OUS) CUSTOMERS WERE NOTIFIED THROUGH AN URGENT FIELD SAFETY NOTICE (UFSN) ACHC24-04.A.OUS.CH, TITLED ¿ADVIA CHEMISTRY URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) LOT 140 QUALITY CONTROL (QC) OUT OF RANGE AND BIASED PATIENT RESULTS¿. CUSTOMERS WERE ADVISED TO DISCONTINUE USE AND DISCARD THE ADVIA UCFP LOT 140. SECTIONS HAS BEEN UPDATED TO REFLECT THE HSC INVESTIGATION. INITIAL MDR 2432235-2024-00128 WAS FILED ON 06-JUN-2024.
Remedial action
- Recall