ALINITY I CEA REAGENT KIT
Received Jun 3, 2024 · Event occurred Jul 4, 2023
Report 3008344661-2024-00065 · MDR key 19454372
Device
Generic name
System, Test, Carcinoembryonic Antigen
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P62-20
Lot number
50043FN00
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, AND DEVICE HISTORY RECORD REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. THE TICKET SEARCH BY LOT INDICATES THAT THE COMPLAINT LOT PERFORMS AS EXPECTED FOR THIS PRODUCT. TICKET TRENDING REVIEW DID NOT IDENTIFY ANY TRENDS. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE COMPLAINT LOT. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I CEA REAGENT LOT 50043FN00 WAS IDENTIFIED.
Description of Event or Problem
THE CUSTOMER OBSERVED DISCREPANT ALINITY I CEA RESULTS GENERATED FOR A PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED: PATIENT 1: ON (B)(6) 2023: RESULT WITH OLD LOT 49146FN00 = 2.0, ON (B)(6) 2023 RESULT WITH NEW LOT 50043FN00 = 3.1. ADDITIONALLY, THE CUSTOMER ALSO NOTED A SHIFT UP IN QUALITY CONTROLS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Description of Event or Problem
THE CUSTOMER OBSERVED DISCREPANT ALINITY I CEA RESULTS GENERATED FOR A PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED: PATIENT 1 (B)(60 2023 RESULT WITH OLD LOT 49146FN00 = 2.0, (B)(6) 2023 RESULT WITH NEW LOT 50043FN00 = 3.1 ADDITIONALLY, THE CUSTOMER ALSO NOTED A SHIFT UP IN QUALITY CONTROLS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.