inforMED
InjuryISH

VARI-FLEX

Received May 27, 2024 · Event occurred Apr 4, 2024

Report 3003764610-2024-00003 · MDR key 19400969

Device

Generic name

Prosthetic Foot

Manufacturer

Ossur Hf

Model number

VFPE8280

Product problems

  • Fracture
  • Insufficient Device Problem Information
  • Fracture
  • Insufficient Device Problem Information

Patient

NA · Unknown

  • Laceration(s)
  • Limb Fracture
  • Laceration(s)
  • Limb Fracture

Narrative

Description of Event or Problem

USER WAS ON A MOTOR CYCLE AND WAS NOT ABLE TO BRAKE. HE FELL AND FRACTURED HIS RIGHT HAND AND GOT A LACERATION ON THE RIGHT KNEE.

Description of Event or Problem

USER WAS DRIVING A MOTOR VEHICLE AND WAS NOT ABLE TO BRAKE. HE GOT IN AN ACCIDENT AND FRACTURED HIS RIGHT HAND AND GOT A LACERATION ON THE RIGHT KNEE.

Additional Manufacturer Narrative

THE TOP CARBON BLADES SNAPPED AT THE ADAPTER; THE RIGHT BLADE SHOWED NO DELAMINATION, AND THE LEFT BLADE HAD MINOR DELAMINATION. IT IS CONCLUDED THAT THE INJURIES AND PRODUCT FAILURE WERE A RESULT OF THE CRASH. THE FAILURE CLOSELY RESEMBLES THE RESULTS OF THE DROP TESTS INVOLVING SUDDEN IMPACT OF EXTREME FORCE FOCUSED ON THE TOE AREA. THE PRODUCT HAD BEEN IN USE FOR ONLY 11 DAYS BEFORE THE INCIDENT. THE LIKELIHOOD OF THIS TYPE OF FAILURE LEADING TO A HAZARDOUS EVENT RESULTING IN A SERIOUS INJURY IS CONSIDERED REMOTE. FURTHER ACTIONS ARE NOT CONSIDERED WARRANTED.