KETONE
Received May 16, 2024 · Event occurred Apr 22, 2024
Report 1000113657-2024-00220 · MDR key 19336853
Device
Generic name
Nitroprusside, Ketones (urinary, Non-quant.)
Manufacturer
Trividia Health IncModel number
STRIP, WALGREENSKETONE 50CT #496636Lot number
AA669
Product problems
- Product Quality Problem
- Product Quality Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INTERNAL REPORT REFERENCE NUMBER: (B)(4). KETONE TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-020: USER'S TEST STRIP HAD POOR STORAGE. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.
Description of Event or Problem
CONSUMER REPORTED COMPLAINT FOR THE TRUEPLUS KETONE TEST STRIPS. COMPLAINT WAS INITIALLY REPORTED BY E-MAIL AND CUSTOMER WAS CONTACTED VIA TELEPHONE. THE CUSTOMER REPORTED COMPLAINT FOR PHYSICAL DEFECT OF TEST STRIPS. CUSTOMER STATED THAT SHE HAD PURCHASED THE KETONE TEST STRIPS A FEW WEEKS PRIOR AND WAS JUST USING THEM FOR THE FIRST TIME TODAY. CUSTOMER STATED WHEN SHE OPENED THE VIAL THE TEST STRIPS LOOKED NORMAL COLOR, BUT WHEN SHE USED THEM- THEY TURNED BROWN. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS AND MEDICAL ATTENTION RELATED TO THE USE OF THE PRODUCT WAS NOT REPORTED.
Remedial action
- Replace