UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM

Received Dec 18, 2010 · Event occurred Nov 18, 2010

Report 2050012-2010-01604 · MDR key 1932490

Device

Generic name

Clinical Chemstry Analyzer

Model number

DXC800 PRO

Catalog number

A11812

Lot number

NA

Product problems

  • Low Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUSLY LOW PHOSPHOROUS (PHOSM) RESULTS GENERATED BY UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM FOR 43 PATIENTS. FOURTEEN RESULTS WERE REPORTED OUT OF THE LABORATORY AND AMENDED BEFORE PATIENT TREATMENT WAS IMPACTED.

Additional Manufacturer Narrative

QC AND CALIBRATION RESULTS WERE ACCEPTABLE PRIOR TO THE EVENT. THE INSTRUMENT WOULD NOT CALIBRATE FOR PHOSM AFTER THE EVENT. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE PHOSM MODULE.

Remedial action

  • Inspection
  • Replace