Received May 9, 2024 · Event occurred Apr 9, 2024

Report 8030229-2024-04287 · MDR key 19282102

Device

Generic name

Electroencephalograph

Model number

EEG-1200A

Catalog number

EEG-1200A

Lot number

NA

Product problems

  • Output Problem
  • Output Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DETAILS OF THE COMPLAINT: THE CUSTOMER REPORTED THAT THE EXAMS LISTED HAVE INCORRECT MEDICAL RECORD NUMBERS. THEY CAN EDIT THE MEDICAL RECORDS NUMBER, AND IT CHANGES, AND THEN IN 24 HOURS IT CHANGES BACK. IT APPEARS THAT WHEN YOU CLICK TO CHECK FOR ACCURACY THE CUSTOMER DOES NOT HAVE RIGHTS TO THE FOLDER TO MAKE THE CHANGES. NOT IN PATIENT USE. INVESTIGATION CONCLUSION: THE NIHON KOHDEN NEUROLOGY ENGINEERING SERVICE TEAM (NK NES) CHECKED THE CUSTOMER DATABASE AND FOUND THAT A DATABASE REGISTRATION FILE WAS RECREATED AND NOT GETTING DELETED NORMALLY AS PART OF THE SUB TO MAIN DATABASE PROCESS WHENEVER NEUROWORKBENCH (NWB) WAS LAUNCHED. NK NES USED AN INSTALLATION TOOL TO KEEP THE EXTRA FILE FROM BEING CREATED AS A WORKAROUND SOLUTION. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED FROM NK NES' REVIEW OF THE CUSTOMER DATABASE, BUT A TEMPORARY SOLUTION WAS CREATED FOR THE CUSTOMER. POSSIBLE CAUSE MAY BE A SOFTWARE DEFICIENCY WITH THE DATABASE MANAGEMENT TOOLS USED WITH NWB. A REVIEW OF THE CUSTOMER'S COMPLAINT HISTORY DOES NOT SHOW RECURRENCE OR OTHER SIMILAR COMPLAINTS. THE FOLLOWING FIELDS CONTAINS NO INFORMATION (NI), AS ATTEMPTS TO OBTAIN INFORMATION WERE MADE, BUT THE INFORMATION WAS NOT PROVIDED. D4 SERIAL NUMBER H4 DEVICE MANUFACTURER DATE ATTEMPT #1: (B)(6) 2024 EMAILED CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL ITEMS UNDER THE NO INFORMATION SECTION. NO REPLY WAS RECEIVED. ATTEMPT #2 (B)(6) 2024 EMAILED CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL ITEMS UNDER THE NO INFORMATION SECTION. THE CUSTOMER RESPONDED BACK WITH COMPLAINT DETAILS, BUT THEY DID NOT PROVIDE THE PATIENT AND DEVICE INFORMATION AS REQUESTED. EEG REVIEW UNIT. MODEL: EEG-1200 REVIEW UNIT SN: NA. CORRECTED INFORMATION: D1 BRAND NAME WAS CORRECTED FROM EEG-1200A TO NEUROFAX

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE EXAMS LISTED HAVE INCORRECT MEDICAL RECORD NUMBERS. THEY CAN EDIT THE MEDICAL RECORDS NUMBER, AND IT CHANGES, AND THEN IN 24 HOURS IT CHANGES BACK. IT APPEARS THAT WHEN YOU CLICK TO CHECK FOR ACCURACY THE CUSTOMER DOES NOT HAVE RIGHTS TO THE FOLDER TO MAKE THE CHANGES. NOT IN PATIENT USE.

Additional Manufacturer Narrative

DETAILS OF THE COMPLAINT: THE CUSTOMER REPORTED THAT THE EXAMS LISTED HAVE INCORRECT MEDICAL RECORD NUMBERS. THEY CAN EDIT THE MEDICAL RECORDS NUMBER, AND IT CHANGES, AND THEN IN 24 HOURS IT CHANGES BACK. IT APPEARS THAT WHEN YOU CLICK TO CHECK FOR ACCURACY THE CUSTOMER DOES NOT HAVE RIGHTS TO THE FOLDER TO MAKE THE CHANGES. NOT IN PATIENT USE. INVESTIGATION CONCLUSION: THE NIHON KOHDEN NEUROLOGY ENGINEERING SERVICE TEAM (NK NES) CHECKED THE CUSTOMER DATABASE AND FOUND THAT A DATABASE REGISTRATION FILE WAS RECREATED AND NOT GETTING DELETED NORMALLY AS PART OF THE SUB TO MAIN DATABASE PROCESS WHENEVER NEUROWORKBENCH (NWB) WAS LAUNCHED. NK NES USED AN INSTALLATION TOOL TO KEEP THE EXTRA FILE FROM BEING CREATED AS A WORKAROUND SOLUTION. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED FROM NK NES' REVIEW OF THE CUSTOMER DATABASE, BUT A TEMPORARY SOLUTION WAS CREATED FOR THE CUSTOMER. POSSIBLE CAUSE MAY BE A SOFTWARE DEFICIENCY WITH THE DATABASE MANAGEMENT TOOLS USED WITH NWB. A REVIEW OF THE CUSTOMER'S COMPLAINT HISTORY DOES NOT SHOW RECURRENCE OR OTHER SIMILAR COMPLAINTS. THE FOLLOWING FIELDS CONTAINS NO INFORMATION (NI), AS ATTEMPTS TO OBTAIN INFORMATION WERE MADE, BUT THE INFORMATION WAS NOT PROVIDED. D4 SERIAL NUMBER. H4 DEVICE MANUFACTURER DATE. ATTEMPT #1: (B)(6) 2024 EMAILED CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL ITEMS UNDER THE NO INFORMATION SECTION. NO REPLY WAS RECEIVED. ATTEMPT #2: (B)(6) 2024 EMAILED CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL ITEMS UNDER THE NO INFORMATION SECTION. THE CUSTOMER RESPONDED BACK WITH COMPLAINT DETAILS, BUT THEY DID NOT PROVIDE THE PATIENT AND DEVICE INFORMATION AS REQUESTED. EEG REVIEW UNIT: MODEL: EEG-1200 REVIEW UNIT: SN: NA. CORRECTED INFORMATION: D1 BRAND NAME WAS CORRECTED FROM EEG-1200A TO NEUROFAX D4 UNIQUE IDENTIFIER (UDI) CORRECTED FROM (B)(4).

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE EXAMS LISTED HAVE INCORRECT MEDICAL RECORD NUMBERS. THEY CAN EDIT THE MEDICAL RECORDS NUMBER, AND IT CHANGES, AND THEN IN 24 HOURS IT CHANGES BACK. IT APPEARS THAT WHEN YOU CLICK TO CHECK FOR ACCURACY THE CUSTOMER DOES NOT HAVE RIGHTS TO THE FOLDER TO MAKE THE CHANGES. NOT IN PATIENT USE.

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE EXAMS LISTED HAVE INCORRECT MEDICAL RECORD NUMBERS. THEY CAN EDIT THE MEDICAL RECORDS NUMBER, AND IT CHANGES, AND THEN IN 24 HOURS IT CHANGES BACK. IT APPEARS THAT WHEN YOU CLICK TO CHECK FOR ACCURACY THE CUSTOMER DOES NOT HAVE RIGHTS TO THE FOLDER TO MAKE THE CHANGES. NOT IN PATIENT USE.

Additional Manufacturer Narrative

THE CUSTOMER REPORTED THAT THE EXAMS LISTED HAVE INCORRECT MEDICAL RECORD NUMBERS. THEY CAN EDIT THE MEDICAL RECORDS NUMBER, AND IT CHANGES, AND THEN IN 24 HOURS IT CHANGES BACK. IT APPEARS THAT WHEN YOU CLICK TO CHECK FOR ACCURACY THE CUSTOMER DOES NOT HAVE RIGHTS TO THE FOLDER TO MAKE THE CHANGES. NOT IN PATIENT USE. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. THE FOLLOWING FIELDS CONTAINS NO INFORMATION (NI), AS ATTEMPTS TO OBTAIN INFORMATION WERE MADE, BUT THE INFORMATION WAS NOT PROVIDED. D4 SERIAL NUMBER H4 DEVICE MANUFACTURER DATE ATTEMPT #1 (B)(6)/2024 EMAILED CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL ITEMS UNDER THE NO INFORMATION SECTION. NO REPLY WAS RECEIVED. ATTEMPT #2 (B)(6)2024 EMAILED CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL ITEMS UNDER THE NO INFORMATION SECTION. THE CUSTOMER RESPONDED BACK WITH COMPLAINT DETAILS, BUT THEY DID NOT PROVIDE THE PATIENT AND DEVICE INFORMATION AS REQUESTED. EEG REVIEW UNIT MODEL: EEG-1200 REVIEW UNIT SN: NA.