WALRUS 087 BALLOON GUIDE CATHETER
Received May 1, 2024 · Event occurred Jun 25, 2023
Report 3015614180-2024-00005 · MDR key 19226933
Device
Generic name
Catheter Introducer
Manufacturer
Q'apel Medical Inc.Model number
087 BALLOON GUIDE CATHETERCatalog number
BG8087-095
Lot number
FG220817J-02
Product problems
- Insufficient Device Problem Information
- Insufficient Device Problem Information
Patient
NA · Unknown
- Vascular Dissection
- Vascular Dissection
Narrative
Additional Manufacturer Narrative
CONCLUSION: UNABLE TO DETERMINE ROOT CAUSE. THE INFORMATION PROVIDED TO Q'APEL DOES NOT STATE THAT A PRODUCT MALFUNCTION OCCURRED. AN ADVERSE EVENT APPEARS TO HAVE OCCURRED BUT THERE IS NOT ENOUGH INFORMATION AVAILABLE TO CLASSIFY THE DEVICE PROBLEM. FG 00100-02, LOT FG220817J-02 WAS NOT RETURNED TO Q'APEL.
Description of Event or Problem
RETROACTIVE REPORTING TO THE FDA OF AN INCIDENT THAT OCCURED IN THE EU MARKET (GERMANY) ON 06/25/2023. Q'APEL MEDICAL BECAME AWARE OF THE INCIDENT ON 10-26-2023. Q'APEL MEDICAL FILED AN MIR REPORT ON 11/10/2023. REASON FOR COMPLAINT: DISSECTION AFTER USING WALRUS. IN AN EMAIL FROM (B)(6) DATED 10/31/2023, 3 CASES FROM THE SAME HOSPITAL AND ATTENDING PHYSICIAN WERE REPORTED TO Q'APEL QUALITY. (B)(6) WILL REPORT FINDINGS FOR CASE 3 AS DOCUMENTED IN EMAILS AND PHYSICIAN'S REPORTING NOTES. "DISSECTION AFTER USING WALRUS" THROMBECTOMY OF M2 SEGMENT TYPE OF INJURY AND LOCATION: DISSECTION IN 2 PARTS OF ICA ( PETROUS AND CAVERNOUS). M2 CLOSURE ON THE RIGHT. PLACEMENT IN THE CAVERNOUS SEGMENT. AFTER THROMBECTOMY KL. DISSECTION UNDER THE ENTRANCE TO THE ROCK BONE AND IN THE HORIZONTAL SECTION OF THE PARS PETROSA. AFTER 1 D CLOSURE OF THE ACI: 2 DISSECTIONS, ONE AT THE ENTRANCE TO THE PELVIC LEG, A SECOND CAVERNOUS WHERE THE BALLOON WAS INFLATED. RECANALIZATION WITH 2 STENTS.