ADVIA CENTAUR CEA
Received May 1, 2024 · Event occurred Apr 10, 2024
Report 1219913-2024-00302 · MDR key 19221923
Device
Generic name
System, Test, Carcinoembryonic Antigen
Manufacturer
Siemens Healthcare Diagnostics Inc.Model number
N/ACatalog number
10309977
Lot number
219
Product problems
- High Test Results
- High Test Results
Patient
48 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A UNITED STATES (US) CUSTOMER REPORTED OBSERVATION OF AN ELEVATED CARCINOEMBRYONIC ANTIGEN (CEA) RESULT WHICH WAS DISCORDANT RELATIVE TO REPEAT TESTING ON THE SAME ADVIA CENTAUR CP ANALYZER. SIEMENS HEALTHCARE DIAGNOSTICS HAS CONCLUDED THE INVESTIGATION OF THE EVENT. SIEMENS HAS EVALUATED THE INSTRUMENT DATA AND INFORMATION PROVIDED BY THE CUSTOMER. THE SAMPLE WAS CHECKED FOR INTEGRITY AND NO FIBRIN OR HIL INTERFERENCE WAS PRESENT. REAGENT OR INSTRUMENT ISSUES WERE RULED OUT BASED ON QUALITY CONTROL (QC) RECOVERY WITHIN ACCEPTABLE RANGES, SAMPLE RESULTS WERE REPEATABLE, AND NO ISSUES WERE REPORTED WITH OTHER PATIENT SAMPLES. BASED ON THE AVAILABLE INFORMATION, THE CAUSE OF THE DISCORDANT RESULT COULD NOT BE DETERMINED. A PRODUCT PROBLEM HAS NOT BEEN IDENTIFIED.
Description of Event or Problem
THE CUSTOMER REPORTS OBSERVATION OF AN ELEVATED ADVIA CENTAUR CARCINOEMBRYONIC ANTIGEN (CEA) RESULT FOR ONE PATIENT WHICH WAS DISCORDANT RELATIVE TO REPEAT TESTING. THE INITIAL ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN. THE SAME SAMPLE WAS RETESTED ON THE ORIGINAL ANALYZER AND PRODUCED A LOWER RESULT. THIS LOWER REPEAT RESULT MATCHED THE PATIENT'S LAB HISTORY FOR CEA, AND WAS REPORTED AS CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED CEA RESULT.