Received Apr 30, 2024 · Event occurred Apr 5, 2024

Report 1823260-2024-01312 · MDR key 19212594

Device

Generic name

Alt/sgpt Test

Manufacturer

Roche Diagnostics

Catalog number

08056773190

Lot number

77315801

Product problems

  • Non Reproducible Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SPECIFIC CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. IT WAS FOUND THE CUSTOMER WAS USING THE REAGENTS PAST THE ONBOARD STABILITY.

Additional Manufacturer Narrative

THE COBAS PURE C 303 ANALYTICAL UNIT SERIAL NUMBER WAS (B)(6) . THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COBAS PURE C 303 ANALYTICAL UNIT. THE INITIAL ALT RESULT WAS 2634 U/L WITH A DATA FLAG. THE REPEAT RESULT WITH A DILUTION WAS 1333 U/L WITH A DATA FLAG. THE REPEAT RESULT FROM A NON-ROCHE ANALYZER WAS 410 U/L.