inforMED
MalfunctionIRT

CONAIR

Received Apr 29, 2024 · Event occurred Apr 17, 2024

Report 1222304-2024-00012 · MDR key 19200267

Device

Generic name

Heating Pad

Manufacturer

Conair Llc.

Model number

HP34

Product problems

  • Appropriate Device Problem Term/Code Not Available
  • Appropriate Device Problem Term/Code Not Available

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON (B)(6) 2024 - THE CONSUMER CLAIMS THE PRODUCT MADE A POPPING NOISE AND HAD SPARKS. NO PROPERTY OR INJURIES OCCURED.

Additional Manufacturer Narrative

4/29/2024 - WE HAVE REQUESTED THE DEVICE BE RETURNED TO THE MANUFACTURER FOR AN INVESTIGATION. TO DATE, WE HAVE NOT RECEIVED THE DEVICE.

Description of Event or Problem

4/19/2004 - THE CONSUMER CLAIMS THE PRODUCT MADE A POPPING NOISE AND HAD SPARKS. NO PROPERTY OR INJURIES OCCURED.

Additional Manufacturer Narrative

4/29/2024 - WE HAVE REQUESTED THE DEVICE BE RETURNED TO THE MANUFACTURER FOR AN INVESTIGATION. TO DATE, WE HAVE NOT RECEIVED THE DEVICE. 7/10/2024 - WE HAVE RECEIVED THE DEVICE FROM THE CONSUMER AND COMPLETED AN INVESTIGATION. BELOW IS THE MANUFACTURERS NARRATIVE: THE HEATING PAD MADE A "PUFF" NOISE, POPPING NOISE. HE STATES THE HEATING PAD HAD SOME SPARKS. THIS HAPPENED A WEEK AGO. HE STATES HE CALLED IN BEFORE BUT NOT DOCUMENTED. THERE WAS NO PROPERTY DAMAGE NOR PERSONAL INJURY. NO TESTING POSSIBLE. HEATED THROW BLANKET SAMPLE WAS IN A NON WORKING CONDITION. AS UNIT COULD NOT BE TESTED, CONTROLLER WAS TAKEN APART AND A COMPONENT WAS FOUND TO HAVE FAILED. THE COMPONENT FAILURE WAS THE MOV, WHICH IS DESIGNED TO ABSORB VOLTAGE SPIKES DUE TO THE POWER SUPPLY. THE CONDITION OF THE COMPONENT WOULD INDICATE IT PREVENTED A LARGER ELECTRICAL FAILURE BY OPENING DUE TOO A VOLTAGE SPIKE.