MalfunctionDXK

ULTRA ICE PLUS CATHETER

Received Apr 22, 2024 · Event occurred Mar 18, 2024

Report MW5154100 · MDR key 19164022

Device

Generic name

Echocardiograph

Model number

M00499120

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

DURING PAROXYSMAL ATRIAL FIBRILLATION PROCEDURE, AN AIR LEAK OCCURRED. THE SHEATH WAS INSERTED THROUGH THE RIGHT FEMORAL VEIN, THE NON- (B)(6) CATHETER (ULTRA ICE PLUS BOSTON SCIENTIFIC) WAS FLUSHED AND THEN INSERTED THROUGH THE SHEATH WITHOUT ANY NOTED RESISTANCE. WHEN ASPIRATING BLOOD, AIR WAS ASPIRATED INTO THE SYRINGE. PHYSICIAN FELT THAT THE HEMOSTASIS VALVE WAS LOOSE. THE SHEATH WAS REPLACED WITH ANOTHER OF THE SAME MODEL BUT AIR WAS STILL ASPIRATED. THE NON- (B)(6) CATHETER (ULTRA ICE PLUS BOSTON SCIENTIFIC) WAS REPLACED AND THE ISSUE WAS RESOLVED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. THE DEVICE WAS REMOVED BEFORE THE AIR WAS INTRODUCED INTO THE PATIENT. FLUOROSCOPY WAS USED TO CONFIRM THAT NO AIR ENTERED THE PATIENT. THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2).