inforMED
MalfunctionFLE

RENAISSANCE STERILIZER

Received Apr 18, 2024 · Event occurred Mar 18, 2024

Report 3005899764-2024-00016 · MDR key 19130397

Device

Generic name

Sterilizer

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT STEAM WAS LEAKING FROM THEIR RENAISSANCE STERILIZER AND SET OFF THE FIRE ALARM. NO REPORT OF INJURY.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE UNIT AND FOUND THAT THE PRESSURE REGULATING VALVE HAD FAILED WHICH CAUSED THE REPORTED EVENT. THE TECHNICIAN REBUILT THE PRESSURE REGULATING VALVE. THE UNIT WAS TESTED, CONFIRMED TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.