inforMED
MalfunctionPTS

GRASPING FORCEPS

Received Apr 9, 2024 · Event occurred Jan 13, 2023

Report 9614641-2024-00878 · MDR key 19069966

Device

Generic name

Grasping Forceps

Model number

FG-53SX-1

Lot number

K2805-4878

Product problems

  • Device Reprocessing Problem
  • Device Reprocessing Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT A BROWN FOREIGN SUBSTANCE WAS FOUND ATTACHED TO PART OF THE FORCEPS SLIDER OF THE GRASPING FORCEPS. THE ISSUE WAS FOUND WHEN AUTOCLAVE STERILIZATION WAS DONE AFTER A UROLOGY PROCEDURE. THE CUSTOMER SUSPECTED THE FOREIGN MATERIAL MAY HAVE BEEN LEFT UNWASHED DURING CLEANING. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.

Additional Manufacturer Narrative

DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE ADHESIVE AT THE JOINT OF THE CONTROL SECTION NEAR THE SLIDER HAD TURNED BROWN. THERE WAS NO FOREIGN MATTER ATTACHED TO THE SLIDER, BUT THERE WAS FOREIGN MATTER FOUND AROUND THE GRIP. THERE WERE NO ABNORMALITIES SUCH AS BUCKLING IN THE APPEARANCE. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. ALTHOUGH A DEFINITIVE ROOT CAUSE OF THE FOREIGN SUBSTANCE COULD NOT BE DETERMINED, THE FOREIGN SUBSTANCE WAS DETERMINED TO BE DISCOLORED DUE TO DETERIORATION OF THE ADHESIVE. LIKELY FACTORS CAUSING THE ADHESIVE DETERIORATION COULD HAVE BEEN THE FOLLOWING: 1 - RINSING AFTER WASHING WAS INSUFFICIENT, AND THE WASHING LIQUID REMAINED. 2 - THE PRODUCT WAS AUTOCLAVED IN THE STATE DESCRIBED IN (1) ABOVE. 3 - THE ADHESIVE DETERIORATED AND DISCOLORED DUE TO THE COMPONENTS OF THE CLEANING SOLUTION AND THE HEAT OF THE AUTOCLAVE. THE CUSTOMER MAY BE ABLE TO REDUCE AND PREVENT OCCURRENCE OF THE EVENT BY HANDLING DEVICE IN ACCORDANCE WITH THE FOLLOWING IFU : REPROCESS THE INSTRUMENT IMMEDIATELY AFTER USE, FIRST BY IMMERSING IT IN A NEUTRAL, LOW-FOAMING, MEDICAL GRADE DETERGENT SOLUTION, THEN FOLLOWING THE CLEANING AND STERILIZATION PROCEDURES GIVEN IN THIS CHAPTER. FAILURE TO REPROCESS THE INSTRUMENT IMMEDIATELY AFTER USE, OR USING OTHER THAN A MEDICAL-GRADE DETERGENT MAY CAUSE CORROSION AT THE METAL PARTS LIKE THE GRASPING JAWS. THIS COULD IMPAIR THE OPERATION OF THE INSTRUMENT OR CAUSE A PART OF IT TO BREAK AND/OR FALL OFF INSIDE THE PATIENT. IMMERSE THE INSTRUMENT IN DETERGENT SOLUTION IMMEDIATELY AFTER USE. IF THE INSTRUMENT IS NOT CLEANED IMMEDIATELY, IT MAY BE DIFFICULT TO EFFECTIVELY REPROCESS AND THIS COULD RESULT IN REDUCED PERFORMANCE. 1 - WHEN REPROCESSING THE INSTRUMENT FOR THE FIRST TIME AFTER PURCHASE, REMOVE THE FORCEPS CAP FROM THE GRASPING JAWS AND DISPOSE OF IT. 2 - IMMERSE THE ENTIRE INSTRUMENT IN THE DETERGENT SOLUTION FOR THE TIME SPECIFIED IN MANUFACTURER¿S INSTRUCTIONS. IF NO TIME IS SPECIFIED, IMMERSE FOR BETWEEN 5 MINUTES AND 3 HOURS. 3 - REMOVE THE INSTRUMENT FROM THE DETERGENT SOLUTION. ULTRASONIC CLEANING. 1 - IMMERSE THE ENTIRE INSTRUMENT IN THE ULTRASONIC CLEANER CONTAINING DETERGENT SOLUTION. 2 - CLEAN ULTRASONICALLY FOR 30 MINUTES. FOR DETAILS ON OPERATION OF THE ULTRASONIC CLEANER, REFER TO THE INSTRUCTION MANUAL OF THE ULTRASONIC CLEANER. 3 - REMOVE THE INSTRUMENT FROM THE DETERGENT SOLUTION. RINSING - AFTER ULTRASONIC CLEANING, RINSE THE INSTRUMENT THOROUGHLY TO REMOVE RESIDUAL DETERGENT. RESIDUAL DETERGENT SOLUTION COULD CAUSE TISSUE IRRITATION IN THE NEXT PATIENT. 1 - RINSE THE INSTRUMENT UNDER CLEAN RUNNING WATER. 2 - CONFIRM THAT NO DEBRIS IS LEFT ON THE SURFACES OF THE INSTRUMENT. 3 - WIPE THE EXTERIOR OF THE INSTRUMENT WITH A CLEAN, DRY LINT-FREE CLOTH. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AT A LATER DATE, THIS REPORT WILL BE SUPPLEMENTED. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE.