CLEARLINK BLOOD RECIPIENT SET
Received Apr 2, 2024 · Event occurred Mar 1, 2024
Report 1416980-2024-01463 · MDR key 19020890
Device
Generic name
Set, Blood Transfusion
Manufacturer
Baxter Healthcare CorporationModel number
NACatalog number
2C8750
Lot number
DR23K30048
Product problems
- Fluid/Blood Leak
- Material Puncture/Hole
- Fluid/Blood Leak
- Material Puncture/Hole
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT A CLEARLINK SYSTEM Y-TYPE BLOOD/SOLUTION SET LEAKED FROM THE TUBING BELOW THE DRIP CHAMBER. SMALL DROPS OF THE PLATELET PRODUCT DRIPPED OUT THROUGH A SMALL HOLE IN THE TUBING. THE EVENT OCCURRED AFTER THE TUBING WAS PRIMED AS THE ROLLER CLAMPS WERE ALL CLOSED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Additional Manufacturer Narrative
H10: A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE DEVICE WAS NOT RECEIVED FOR EVALUATION; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.