COBAS 6000 C501 MODULE
Received Nov 17, 2010 · Event occurred Nov 1, 2010
Report 1823260-2010-06794 · MDR key 1901377
Device
Generic name
Clinical Chemistry Analyzer
Manufacturer
Roche DiagnosticsModel number
NACatalog number
05036453001
Lot number
NA
Product problems
- Low Test Results
Patient
Not reported
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
THE USER RECEIVED QUESTIONABLE RESULTS FOR TOTAL PROTEIN GENERATION 2 (TP2) ON THE COBAS C501 ANALYZER. THE EVENT INVOLVED 7 PATIENT SAMPLES WHICH GAVE DISCREPANT RESULTS. THE PATIENT SAMPLES WERE REPEATED ON (B)(6) 2010 AND WERE RUN ON A DIFFERENT C501 ANALYZER SERIAL NUMBER (B)(4). PATIENT SAMPLE 1, THE INITIAL RESULT WAS 2.3 G/DL; THE REPEAT RESULT WAS 8.4 G/DL. PATIENT SAMPLE 2, THE INITIAL RESULT WAS 2.3 G/DL; THE REPEAT RESULT WAS 7.7 G/DL. PATIENT SAMPLE 3, THE INITIAL RESULT WAS 2.3 G/DL; THE REPEAT RESULT WAS 6.7 G/DL. PATIENT SAMPLE 4, THE INITIAL RESULT WAS 2.4 G/DL; THE REPEAT RESULT WAS 8.3 G/DL. PATIENT SAMPLE 5, THE INITIAL RESULT WAS 2.4 G/DL; THE REPEAT RESULT WAS 5.4 G/DL. PATIENT SAMPLE 6, THE INITIAL RESULT WAS 2.3 G/DL; THE REPEAT RESULT WAS 5.7 G/DL. PATIENT SAMPLE 7, THE INITIAL RESULT WAS 2.4 G/DL; THE REPEAT RESULT WAS 7.4 G/DL. THE INITIAL RESULTS WERE REPORTED OUTSIDE THE LABORATORY TO THE PHYSICIANS AND REQUIRED CORRECTED REPORTS TO BE GENERATED. THE PATIENTS DID NOT RECEIVE ANY TREATMENT AND WERE NOT AFFECTED BY THIS EVENT. THE REAGENT LOT NUMBER FOR TP2 WAS NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE FOUND AN ISSUE WITH THE GEAR PUMP. HE ADJUSTED GEAR PUMP PRESSURES. PERFORMANCE TESTS WERE RUN WHICH WERE ACCEPTABLE.