ATELLICA CH
Received Apr 1, 2024 · Event occurred Mar 18, 2024
Report 2432235-2024-00062 · MDR key 19010412
Device
Generic name
Urinary/cerebrospinal Fluid Protein (ucfp)
Manufacturer
Siemens Healthcare Diagnostics Inc.Model number
N/ACatalog number
11097543
Lot number
130414
Product problems
- Low Test Results
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
FIVE (5) FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULTS WERE OBTAINED ON AN ATELLICA ® CH 930 ANALYZER. THE FALSELY DEPRESSED PATIENT SAMPLE RESULTS WERE REPORTED TO THE PHYSICIAN AND UNKNOWN IF QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON THE SAME ATELLICA ® CH 930 ANALYZER UTILIZING AN ALTERNATE REAGENT LOT. THE REPEATED SAMPLE RESULTS WITH THE ALTERNATE REAGENT LOT WERE CONSIDERED CORRECT AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Additional Manufacturer Narrative
AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS CUSTOMER CARE CENTER (CCC) REGARDING FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS OBTAINED ON PATIENT SAMPLES WHEN PROCESSED ON AN ATELLICA ® CH 930 ANALYZER. SIEMENS IS INVESTIGATING THE EVENT.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION (11-JULY-2024): SIEMENS HEALTHINEERS HAS CONFIRMED THE POTENTIAL FOR BIASED QUALITY CONTROL (QC) AND PATIENT RESULTS WHEN USING ATELLICA CH URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) LOT 130414 ON THE ATELLICA® CH AND ATELLICA® CI ANALYZERS. SIEMENS¿ INTERNAL INVESTIGATION CONFIRMED WHEN USING LOT 130414, QC MAY RECOVER OUTSIDE OF THE ALLOWABLE CONTROL LIMITS FOR URINE CHEMISTRY AND SPINAL FLUID CONTROL LEVELS. US CUSTOMERS WERE NOTIFIED THROUGH URGENT MEDICAL DEVICE CORRECTION (UMDC LETTER ACHC24-04.A.US), TITLED ¿ATELLICA CH URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) LOT 130414 QUALITY CONTROL (QC) OUT OF RANGE AND BIASED PATIENT RESULTS¿ AND OUS CUSTOMERS WERE IDENTIFIED THROUGH (URGENT FIELD SAFETY NOTICE (UFSN), ACHC24-04.A.OUS) OF THE POTENTIAL FOR BIAS QUALITY CONTROL (QC) AND PATIENT SAMPLE RESULTS WHEN USING THE ATELLICA CH UCFP LOT 130414. THE UMDC AND UFSN PROVIDED INSTRUCTIONS TO CUSTOMERS TO DISCONTINUE USE AND DISCARD UCFP LOT 130414. SECTION H6 HAS BEEN UPDATED TO REFLECT THE HSC INVESTIGATION. INITIAL MDR 2432235-2024-00062 WAS FILED ON 01-APRIL-2024.
Remedial action
- Recall