ELECSYS PROBNP II
Received Mar 29, 2024 · Event occurred Mar 13, 2024
Report 1823260-2024-00962 · MDR key 19006202
Device
Generic name
Natriuretic Peptide Test
Manufacturer
Roche DiagnosticsCatalog number
09315284214
Lot number
73232701
Product problems
- Low Test Results
- Low Readings
- Non Reproducible Results
- Low Test Results
- Low Readings
- Non Reproducible Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
CALIBRATION SIGNALS WERE WITHIN EXPECTATIONS. NO QC WAS PERFORMED ON THE DAY OF THE EVENT. PER PRODUCT LABELING, "CONTROLS FOR THE VARIOUS CONCENTRATION RANGES SHOULD BE RUN INDIVIDUALLY AT LEAST ONCE EVERY 24 HOURS WHEN THE TEST IS IN USE, ONCE PER COBAS E PACK, AND FOLLOWING EACH CALIBRATION". THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
THE SERIAL NUMBER OF THE E801 ANALYZER IS (B)(6) . THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED QUESTIONABLE RESULTS FOR APPROXIMATELY 80 PATIENT SAMPLES TESTED WITH ELECSYS PROBNP II ON A COBAS E 801 MODULE. CONTROLS TESTED ON (B)(6) 2024 WERE ACCEPTABLE AND PATIENT SAMPLES WERE MEASURED ON THE EVENING OF (B)(6) 2024 TO THE MORNING OF (B)(6) 2024. CONTROLS RECOVERED LOW ON THE MORNING OF (B)(6) 2024. THE CONTROLS WERE REPEATED, BUT REMAINED LOW. THE REAGENT PACK WAS THEN REPLACED AND CONTROLS WERE THEN ACCEPTABLE. THE CUSTOMER REPEATED THE PATIENT SAMPLES ON A SECOND ANALYZER. EXAMPLES WERE PROVIDED FOR 3 PATIENT SAMPLES WITH DISCREPANT PROBNP RESULTS. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE FIRST SAMPLE INITIALLY RESULTED IN A PROBNP VALUE OF 1160 PG/ML AND IT REPEATED AS 1520 PG/ML. THE SECOND SAMPLE INITIALLY RESULTED IN A PROBNP VALUE OF 334 PG/ML AND IT REPEATED AS 450 PG/ML. THE THIRD SAMPLE INITIALLY RESULTED IN A PROBNP VALUE OF 263 PG/ML AND IT REPEATED AS 342 PG/ML.