inforMED
MalfunctionGKF

I-STAT CHEM 8+N CARTRIDGE

Received Nov 3, 2010 · Event occurred Oct 11, 2010

Report 2245578-2010-00141 · MDR key 1893966

Device

Generic name

Chem 8+ Cartridge

Catalog number

03P91-25

Lot number

G10170

Product problems

  • High Test Results

Patient

83 YR

  • Death
  • Respiratory Failure

Narrative

Additional Manufacturer Narrative

(B)(4).

Description of Event or Problem

ON (B)(6) 2010, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THE I-STAT CHEM 8+ CARTRIDGE YIELDED A HEMATOCRIT RESULT OF 37% PCV. A SEPARATE SAMPLE, DRAWN AT THE SAME TIME AS THE I-STAT SAMPLE, WAS TESTED ON THE LAB INSTRUMENT AND THE HEMATOCRIT RESULT WAS 15.4% PCV. NO REPEAT LABS PERFORMED. PATIENT WAS NOT TREATED ON EITHER HEMATOCRIT RESULT. PATIENT DIED OF RESPIRATORY FAILURE AT 6:30PM (B)(6) 2010.