inforMED
MalfunctionGKZ

BD MULTITEST¿

Received Mar 15, 2024 · Event occurred Feb 17, 2024

Report 2916837-2024-00022 · MDR key 18916358

Device

Generic name

Counter, Differential Cell

Catalog number

662967

Lot number

69761

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

G.5. PMA / 510(K)#: K090967, K170974. E.12. INITIAL REPORTER FACILITY NAME: (B)(6). H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED YJAY AFTER USING BD MULTITEST¿, ERRONEOUS RESULTS WERE ACQUIRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC , ,CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOT(76010) ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 69761 1. IF THERE WAS ANY CONTAMINATION THAT INVOLVES PATIENT SAMPLE? NO. 2. WERE ERRONEOUS RESULTS OBSERVED ON PATIENT SAMPLES? YES.

Additional Manufacturer Narrative

H6.INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE (NON-SPECIFIC STAINING CREATING UNUSUAL STAINING PATTERNS FOR CD4+CD+8 POPULATIONS) WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: DATA PROVIDED BY CUSTOMER WAS REVIEWED. RETAIN TESTING RESULTS SHOWED THE PRESENCE OF A NON-SPECIFIC STAINING. POTENTIAL CAUSE IS DETERMINED TO BE MANUFACTURING RELATED. ADDITIONAL CONTROLS HAVE BEEN PUT IN PLACE IN THE MANUFACTURING STAGE SPECIFIC TO THE SUPPLIED COMPONENT. CORRECTIVE ACTION WAS INITIATED TO ADDRESS THE ROOT CAUSE AND PREVENTIVE ACTIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem

IT WAS REPORTED THAT AFTER USING BD MULTITEST¿, ERRONEOUS RESULTS WERE ACQUIRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC, ,CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOT (76010) ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 69761. 1. IF THERE WAS ANY CONTAMINATION THAT INVOLVES PATIENT SAMPLE? NO. 2. WERE ERRONEOUS RESULTS OBSERVED ON PATIENT SAMPLES? YES.

Remedial action

  • Recall