inforMED
MalfunctionFLE

EVOLUTION TRANSFER CARRIAGE

Received Mar 15, 2024 · Event occurred Feb 14, 2024

Report 3005899764-2024-00010 · MDR key 18914860

Device

Generic name

Transfer Carriage

Product problems

  • Device Tipped Over
  • Device Tipped Over

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THE EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND THE SHELF WITH STERILE HOT INSTRUMENTS ON IT FELL ONTO THE FLOOR. NO REPORT OF INJURY.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE EVOLUTION TRANSFER CARRIAGE AND FOUND THE SHELVING SUPPORT WAS BENT CAUSING THE SHELVING TO NOT STAY IN PLACE RESULTING IN THE REPORTED EVENT. THE EVOLUTION TRANSFER CARRIAGE WILL BE RETURNED AND REPLACED. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.