Received Mar 14, 2024 · Event occurred Feb 15, 2024

Report 1220948-2024-00063 · MDR key 18907870

Device

Generic name

Tape

Catalog number

1108-20

Lot number

GTF1119

Product problems

  • Human-Device Interface Problem
  • Human-Device Interface Problem

Patient

NA · Unknown

  • Radiation Burn
  • Pain
  • Skin Inflammation/ Irritation
  • Radiation Burn
  • Pain
  • Skin Inflammation/ Irritation

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE VASCUTAPE WAS NOT REMOVED FROM THE PATIENT'S LEG FOR TWO WEEKS AFTER THE PROCEDURE. THE VASCUTAPE PRODUCT IS NOT INTENDED TO BE PLACED ON THE PATIENT FOR MORE THAN 24 HOURS. THE PATIENT UNDERWENT AN MRI WHILE THE PRODUCT WAS STILL ON THE PATIENT. AS A RESULT, THE MARKINGS OF THE TAPE WERE BURNED ONTO THE PATIENT'S LEG DURING THE MRI AND EXCORIATED DOWN TOWARD THE ANKLE CAUSING PEELING AND REDNESS. THE PATIENT REQUIRED TREATMENT FOR WOUND CARE AND ANTIBIOTICS TO PREVENT INFECTION. IT WAS DETERMINED THAT THIS ISSUE IS DUE TO A USER ERROR. THE VASCUTAPE PRODUCT IS NOT INTENDED TO BE LEFT ON THE PATIENT FOR MORE THAN 24 HOURS AND IS NOT SUITABLE FOR USE IN AN MRI PROCEDURE. WHEN THE PRODUCT WAS LEFT ON THE PATIENT, IT CAUSED A REACTION WHEN THE PATIENT WENT THROUGH THE MRI PROCEDURE RESULTING IN THE EVENT. THE LOT HISTORY RECORD FOR THE DEVICE WAS REVIEWED, NO ISSUES WERE FOUND DURING MANUFACTURING OR PACKAGING THAT WOULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. WE HAVE NOT RECEIVED ANY SIMILAR COMPLAINTS RELATED TO THIS LOT NUMBER.

Description of Event or Problem

TAPE WAS NOT REMOVED FROM PATIENT'S LEG FOR TWO WEEKS AFTER PROCEDURE. PATIENT UNDERWENT MRI WITH TAPE INADVERTANTLY ON HIS LEG, RESULTING IN PAIN AND REDNESS, AS WELL AS BURN OF THE TAPE MARKINGS AS AN IMPRINT ON THE PATIENT'S LEG AFTER TAPE WAS FINALLY REMOVED. PATIENT HAD ANGIO PROCEDURE IN IR SUITE ON (B)(6) 2024, FOLLOWED BY LYSIS CASE THE FOLLOWING DAY; ISSUE AROSE WHEN MRI CONDUCTED 2 WEEKS LATER ON (B)(6) 2024 WITH TAPE STILL ON. PATIENT FINALLY REMOVED TAPE FROM HIS LEG (B)(6) 2024, SEVERAL DAYS AFTER MRI, AND OBSERVED THE REDNESS AND MARKINGS. TAPE MEASUREMENT MARKINGS WERE BURNED INTO PATIENT'S LEG DURING MRI; EXCORIATED DOWN TOWARDS THE ANKLE, PEELING AND RED. PATIENT IS BEING TREATED WITH WOUND CARE AND ANTIBIOTICS AND WAS DISCHARTED BUT FOLLOWED UP WITH BY IR SUITE ON (B)(6) 2024.