WAVE SIDE GRASPER FORCEPS
Received Oct 28, 2010 · Event occurred Sep 23, 2010
Report 1418479-2010-00019 · MDR key 1888660
Device
Generic name
Wave Side Grasper Forceps
Manufacturer
Richard Wolf Medical InstrumentsModel number
8383.2937Catalog number
8383.2937
Lot number
1A07
Product problems
- Fracture
Patient
32 YR
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE A SMALL HOLE WAS MADE IN THE TRANSVERSE MESOCOLON JUST TO THE LEFT OF LIGAMENT OF TREITZ. THE SMALL BOWEL WAS THEN DELIVERED THROUGH THIS AND UP INTO THE LESSER SAC. AT THIS POINT, THE PROXIMAL GASTRIC POUCH WAS GRASPED AND ELEVATED. A SMALL GASTROTOMY WAS MADE IN THE LATERAL PORTION OF THE STOMACH. A STAPLER WAS CHOSEN AND UTILIZED AT THIS TIME. THE ANVIL AND TROCAR WERE PLACED INTO THE ABDOMEN. THIS WAS PLACED THROUGH THE OPENING IN THE STOMACH POUCH AND EXITED FROM THE MID PORTION OF THE POUCH. THIS WAS CLOSED USING A STAPLER. WHILE TRYING TO REMOVE THE TROCAR OR THE ANVIL, THE GRASPER INSTRUMENT HAD ONE OF ITS SIDES FRACTURE. THE PIECE WAS REMOVED INTACT. THERE WAS NO INJURY TO THE PATIENT. THERE DID NOT APPEAR TO BE ANY RETAINED PIECES. THE REMOVED PIECE DID FIT NICELY IN WITH THE INSTRUMENT WHERE IT BROKE OFF.
Additional Manufacturer Narrative
RESULTS: OTHER SERVICE ORGANIZATION HAS REPAIRED THIS DEVICE. EVALUATION SUMMARY: VISUAL INSPECTION OF THE FORCEPS WITH BROKEN JAW SHOWED NO CORROSION AT THE POINT OF BREAKAGE. THE BROKEN JAW WAS RETURNED WITH THE FORCEPS. THERE WAS NO EVIDENCE OF CORROSION ON THE JAW. THE FORCEPS ARE APPROXIMATELY (B)(6). THERE IS A DATE ETCHED ON THE HANDLE 1/10 R THAT INDICATES A REPAIR BY A REPAIR SERVICE OTHER THAN RICHARD WOLF MEDICAL INSTRUMENTS. OUR SERVICE RECORDS FOR THIS FACILITY DID NOT SHOW ANY REPAIRS DONE WITH THIS LOT NUMBER 1A07. FOLLOW-UP WITH THE FACILITY CONTACT PERSON DID INDICATE THAT SOMETIMES THE SURGEON WILL TORQUE THE FORCEPS. THE MANUFACTURER WILL BE INVESTIGATING THE FORCEPS AND A FOLLOW-UP WILL BE DONE SHOULD NEW DEVELOPMENTS OCCUR. CAUSE OF EVENT: NO CONCLUSION CAN BE DRAWN.