inforMED
MalfunctionOQU

SONARMED

Received Mar 11, 2024 · Event occurred Dec 20, 2023

Report 1282497-2024-00029 · MDR key 18883100

Device

Generic name

Airway Monitoring System

Manufacturer

Sonarmed, Inc.

Model number

AW-S030

Catalog number

AW-S030

Product problems

  • Defective Component
  • Defective Component

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

SECTION H9: 806 NUMBER 3009023584-03-08-2024-01-R SECTION H6 EVALUATION CODES UPDATED DUE TO COMPLETION OF INVESTIGATION RESULT - ADD ADDITIONAL CODE CONCLUSION - REMOVE D15 AND ADD D03 COMPONENT - REMOVE G07003 AND ADD G03012 IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THE CLINICIAN HAD AN ISSUE WITH PASSING A 8FRENCH CATHETER THROUGH THIS 3.0 SONARMED SENSOR. ALTHOUGH REQUESTED, INFORMATION PERTAINING TO PATIENT INTERVENTION AND OUTCOME HAS NOT BEEN PROVIDED. NO PHYSICAL PRODUCT OR PHOTOS WERE RETURNED FOR ANALYSIS. ALTHOUGH NO PRODUCT WAS RETURNED, THE LIKELY CAUSE OF THE ISSUE WAS ISOLATED TO AN OBSTRUCTION IN THE SENSOR. AN EXISTING CORRECTIVE ACTION ADDRESSES THE ISSUE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THE CLINICIAN HAD AN ISSUE WITH PASSING A 8FRENCH CATHETER THROUGH THIS 3.0 SONARMED SENSOR. THE CATHETER FIT BUT WAS SO TIGHT AND TOOK SO LONG TO ADVANCE IT OCCLUDED THE ENDOTRACHEAL TUBE (ETT) THAT THE BABY STARTED TO ¿BRADY¿. ALTHOUGH REQUESTED, INFORMATION PERTAINING TO PATIENT INTERVENTION AND OUTCOME HAS NOT BEEN PROVIDED.

Additional Manufacturer Narrative

H3: THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE INVESTIGATION IS CURRENTLY IN PROGRESS. THERE IS AN EXISTING PREVIOUS INTERNAL INVESTIGATION REGARDING THE REPORTED ISSUE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Remedial action

  • Recall