CUSTOM ANGIOGRAPHIC KIT
Received Mar 7, 2024 · Event occurred Feb 21, 2024
Report 1721504-2024-00013 · MDR key 18857341
Device
Generic name
Intravenous Extension Tubing Set -
Manufacturer
Merit Medical Systems Inc.Catalog number
K08-MP5411B
Lot number
H2757517
Product problems
- Unsealed Device Packaging
- Unsealed Device Packaging
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE ACCOUNT ALLEGES THAT THERE WAS A DEFECT IN THE PACKAGING SEAL RESULTING IN INCOMPLETE STERILIZATION OF THE CONTENTS. THIS WAS IDENTIFIED DURING AN INITIAL INSPECTION OF THE RECEIVED PRODUCT. THE DEVICE WAS NOT SENT TO A USER FACILITY.
Additional Manufacturer Narrative
THE SUSPECT MEDICAL DEVICE HAS BEEN RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND MICROSCOPICALLY INVESTIGATED. THE PACKAGING WAS TESTED, AND THE STERILITY WAS FOUND TO BE COMPROMISED. THE COMPLAINT IS CONFIRMED. THE ROOT CAUSE IS ATTRIBUTED TO THE MANUFACTURING PROCESS. THE DEVICE HISTORY RECORD WAS REVIEWED, AND NO EXCEPTION DOCUMENTS WERE FOUND. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE IDENTIFIED. THE DEVICE WAS NOT SENT TO THE USER FACILITY.