V-PRO MAX STERILIZER
Received Mar 7, 2024 · Event occurred Feb 8, 2024
Report 3005899764-2024-00009 · MDR key 18856945
Device
Generic name
Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Product problems
- Use of Device Problem
- Use of Device Problem
Patient
NA · Unknown
- Burn(s)
- Burn(s)
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE STERILIZER AND FOUND THE UNIT TO BE OPERATING PROPERLY. NO ISSUES WITH THE FUNCTION OR OPERATION OF THE STERILIZER WERE IDENTIFIED AND THE STERILIZER WAS RETURNED TO SERVICE. THE REPORTED BURN IS ATTRIBUTED TO THE EMPLOYEE NOT WEARING PROPER PPE, SPECIFICALLY GLOVES, WHILE OPERATING THEIR V-PRO MAX STERILIZER. THE V-PRO MAX STERILIZER OPERATOR MANUAL (1-2) STATES, "DANGER - CHEMICAL INJURY HAZARD: ANY VISIBLE LIQUIDS IN THE CHAMBER OR ON THE LOAD MUST BE TREATED AS CONCENTRATED HYDROGEN PEROXIDE. OBSERVE ALL HYDROGEN PEROXIDE HANDLING PRECAUTIONS. WHEN HANDLING HYDROGEN PEROXIDE, WEAR APPROPRIATE PERSONAL PROTECTIVE EQUIPMENT." THE V-PRO MAX OPERATOR MANUAL (1-2) STATES, "ANSI/AAMI ST58, 2013, RECOMMENDS USING CHEMICAL-RESISTANT GLOVES WHEN USING THE STERILIZATION UNIT." ADDITIONALLY, USER FACILITY PERSONNEL SHOULD ENSURE ALL INSTRUMENTS ARE PROPERLY DRY PRIOR TO PLACEMENT IN THE V-PRO MAX STERILIZER. THE V-PRO MAX OPERATOR MANUAL, (A-1) STATES, "DRY ALL ITEMS THOROUGHLY. ENSURE ALL MOISTURE IS REMOVED FROM ALL INTERNAL PARTS (INCLUDING LUMENS). IF NOT, RESIDUAL HYDROGEN PEROXIDE MAY REMAIN AT CYCLE COMPLETION AND/OR A CYCLE ABORT OCCURS. ONLY DRY ITEMS ARE TO BE PLACED IN STERILIZATION UNIT." THE TECHNICIAN COUNSELED USER FACILITY PERSONNEL ON THE PROPER USE AND OPERATION OF THEIR V-PRO MAX STERILIZER, SPECIFICALLY THE IMPORTANCE OF WEARING PROPER PPE AND PROPERLY DRYING INSTRUMENTS. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT AN EMPLOYEE EXPERIENCED A BURN ON THEIR HANDS WHILE HANDLING ITEMS THAT WERE PROCESSED IN A V-PRO MAX STERILIZER. THE EMPLOYEE WAS TAKEN TO THE EMERGENCY ROOM AND RECEIVED MEDICAL TREATMENT (BANDAGE).