AMIS MOBILE LEG POSITIONER 2.0 FRAME
Received Mar 7, 2024 · Event occurred Feb 13, 2024
Report 3005180920-2024-00102 · MDR key 18854200
Device
Generic name
Hip Instrument
Manufacturer
Medacta International SaCatalog number
01.15.10.0190
Lot number
1213764
Product problems
- Insufficient Device Problem Information
- Insufficient Device Problem Information
Patient
96 YR · Female
- Bone Fracture(s)
- Bone Fracture(s)
Narrative
Description of Event or Problem
BONE FRACTURE OCCURRED DURING THE EXTERNAL ROTATION OF THE LEG WITH THE AMIS LEG POSITIONER. NO ADDITIONAL TREATMENT WAS PERFORMED AND THE SURGERY WAS COMPLETED SUCCESSFULLY. (B)(6) YEAR-OLD PATIENT WITH POOR BONE QUALITY. IT IS UNKNOWN WHETHER OR NOT THE SURGEON WILL FIX THE FRACTURE.
Additional Manufacturer Narrative
BATCH REVIEW PERFORMED ON 21 FEBR 2024. LOT 1213764: (B)(4) ITEM MANUFACTURED AND RELEASED ON 23-JUL-2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DEPARTMENT DURING PHA SURGERY ON A 96 YEAR OLD PATIENT, THE TIBIA FRACTURED ON THE DISTAL PART. ACCORDING TO REPORT, THE FRACTURE OCCURED DURING THE EXTERNAL ROTATION OF THE LEG WITH THE AMIS LEG POSITIONER. IT MAY BE POSSIBLE THAT DUE TO THE LOW BONE QUALITY OF THE PATIENT AND THE SMALL OVERALL DIMENSION OF THE TIBIA THE APPLIED TORTIONAL FORCE RESULTED EXCESSIVE.