MalfunctionJIL

COMBUR 7 TEST

Received Mar 6, 2024 · Event occurred Feb 12, 2024

Report 1823260-2024-00679 · MDR key 18850827

Device

Generic name

Urine Test Strips

Manufacturer

Roche Diagnostics

Catalog number

11008552191

Lot number

67817803

Product problems

  • False Negative Result
  • Incorrect Measurement
  • Incorrect, Inadequate or Imprecise Result or Readings
  • False Negative Result
  • Incorrect Measurement
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER ALLEGED FALSE NEGATIVE LEUKOCYTE RESULTS WHILE USING THE COMBUR 7 TEST STRIPS. THE CUSTOMER TESTED THE PATIENT SAMPLE MULTIPLE TIMES AND ALWAYS RECEIVED A NEGATIVE RESULT. THE PATIENT WAS TESTED USING ANOTHER MANUFACTURER TEST STRIP AND THE RESULT WAS POSITIVE. THIS WAS THE EXPECTED RESULT BASED ON THE PATIENT'S SYMPTOMS AND KNOWN CONDITION.

Additional Manufacturer Narrative

THE RETENTION MATERIAL OF LOT 67817800 WAS TESTED BY VISUAL READING WITH NATIVE-URINE AND A LEUKOCYTES-DILUTION-SERIES AND THE RESULTS WERE ACCEPTABLE. THE RETENTION MATERIAL SHOWED NO ABNORMAL RESULTS AND FULFILLED THE REQUIREMENTS. THE TEST STRIPS WERE REQUESTED FROM THE CUSTOMER FOR FURTHER INVESTIGATION. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative

THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.