PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE
Received Mar 6, 2024 · Event occurred Feb 20, 2024
Report 8030965-2024-03316 · MDR key 18850102
Device
Product problems
- Break
- Break
Patient
NA · Male
- Insufficient Information
- Insufficient Information
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. DEVICE HISTORY REVIEW (DHR): PART# 399.480, LOT # 1808869, MANUFACTURING SITE: SYNTHES GMBH, SUPPLIER : NA, RELEASE TO WAREHOUSE DATE: 08 FEB 2008, EXPIRATION DATE: NA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D9: COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. H3, H4, H6: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2024 DURING THE SURGICAL PROCEDURE, THE DISSECTOR IN QUESTION SPLIT INTO TWO PIECES DURING DISSECTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER DISSECTOR. ALL GENERATED FRAGMENTS WERE REMOVED FROM THE PATIENT WITHOUT ADDITIONAL MEDICAL INTERVENTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D9 CORRECTED: E4 H3, H4. H6: THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. THE J&J MEDTECH ORTHOPAEDICS TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE HANDLE OF THE PERIOST-ELEV SLIGHTLY CURV-BLADE STRAIGH WAS FOUND BROKEN ACROSS THE PIN AND ONLY TWO FRAGMENTS OF THE HANDLE WERE RECEIVED. THE OVERALL SURFACE OF THE DEVICE SHOW SIGNS OF WEAR DUE THE USE OVER THE YEARS. A DIMENSIONAL INSPECTION FOR THE PERIOST-ELEV SLIGHTLY CURV-BLADE STRAIGH WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE PERIOST-ELEV SLIGHTLY CURV-BLADE STRAIGH WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW (DHR): PART# 399.480 LOT # 1808869 MANUFACTURING SITE: SYNTHES GMBH SUPPLIER: NA RELEASE TO WAREHOUSE DATE: 08 FEB 2008 EXPIRATION DATE: NA A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.