POWERCEL POWER PROCESSOR

Received Oct 27, 2010 · Event occurred Sep 1, 2010

Report 2050012-2010-01077 · MDR key 1884347

Device

Generic name

Sample Processing System

Model number

NA

Catalog number

A36587

Lot number

NA

Product problems

  • False Positive Result

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE CUSTOMER IS USING SST II PLASTIC TUBES. THE SAMPLES WERE SPUN AT 1960G FOR 10 MINUTES. NO INFORMATION REGARDING PATIENTS AND THE SYSTEM WAS PROVIDED. SERVICE INFORMATION WAS REQUESTED, BUT WAS NOT SUPPLIED. THE ROOT CAUSE FOR THE EVENT HAS NOT BEEN DETERMINED TO DATE.

Description of Event or Problem

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO INCREASED NUMBER OF FALSE POSITIVE HEPATITIS B VIRUS SURFACE ANTIGEN (HBSAG) RESULTS OBTAINED FROM UNICEL DXI 800 ACCESS IMMUNOASSAY ANALYZER AFTER HAVING THE ANALYZER CONNECTED TO POWERCEL POWER PROCESSOR. NO PATIENT TREATMENT WAS AFFECTED BECAUSE ALL POSITIVE RESULTS WERE RUN ON CONFIRMATORY TESTS.