inforMED
MalfunctionPRN

NIMBUS II PLUS AMBULATORY INFUSION PUMP

Received Mar 1, 2024 · Event occurred Feb 23, 2024

Report 3011581906-2024-00212 · MDR key 18820076

Device

Generic name

Ambulatory Infusion Pump

Manufacturer

Infutronix, Llc

Model number

NIMBUS II PLUS

Catalog number

NIMBUS II PLUS

Lot number

2208092640

Product problems

  • No Audible Alarm
  • Protective Measures Problem
  • No Audible Alarm
  • Protective Measures Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ANALYSIS OF THE RETURNED DEVICE WAS COMPLETED ON 3/26/2024. THE RESULTS ARE BELOW: DURING FUNCTIONAL TESTING, THE REPORTED PROBLEM WAS DUPLICATED. THE PUMP WAS POWERED ON, AND DURING POST, THE PUMP DID NOT MAKE ANY SOUND. EACH KEY WAS PRESSED, AND THE PUMP DID NOT MAKE ANY SOUND AT ANY POINT.

Description of Event or Problem

ON (B)(6) 2024, INFUTRONIX RECEIVED A REPORT THAT A PUMP HAD "SPEAKER MODULE NO SOUND," CAUSING LOSS OF INFUSION PARAMETERS. THE INFUSION CANNOT RESUME WITHOUT CAUSING DELAY IN TREATMENT. REQUESTED DEVICE TO BE RETURNED.

Additional Manufacturer Narrative

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. THE PUMP PASSED ALL PREVIOUS TESTS. COMPLAINT DATA WAS REVIEWED, THERE ARE NO PREVIOUS COMPLAINTS ON THIS DEVICE. DEVICE RETURN REQUESTED. THIS MDR WILL BE REOPENED AND UPDATED IN THE EVENT THE DEVICE INVOLVED OR ADDITIONAL INFORMATION BECOMES AVAILABLE.