inforMED
MalfunctionIOY

SUPPORT ARM 176

Received Feb 27, 2024 · Event occurred Feb 14, 2024

Report 8010042-2024-00374 · MDR key 18787821

Device

Generic name

Support, Arm

Model number

SUPPORT ARM 176

Catalog number

6405976

Product problems

  • Fracture
  • Fracture

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

MANUFACTURER'S REF. #: (B)(4).

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE VENTILATOR'S SUPPORT ARM WAS BROKEN. THE VENTILATOR WAS INVESTIGATED ON SITE BY OUR FIELD SERVICE ENGINEER AND A PHYSICAL DAMAGE TO THE CLAP ATTACHMENT OF THE SUPPORT ARM WAS CONFIRMED. DAMAGED CLAMP ATTACHMENT ALSO CONFIRMED IN THE ATTACHED PHOTO. NO INFORMATION PROVIDED BY CUSTOMER ON WHAT CAUSED THE CLAMP ATTACHMENT TO BREAK. ISSUE RESOLVED BY REPLACING THE SUPPORT ARM AND THE VENTILATOR WAS HANDED OVER TO CUSTOMER IN WORKING CONDITION. THE ROOT CAUSE CAN NOT BE ESTABLISHED BY THIS INVESTIGATION.

Description of Event or Problem

IT WAS REPORTED THAT THE VENTILATOR'S SUPPORT ARM WAS BROKEN. THERE WAS NO PATIENT HARM REPORTED. MANUFACTURER'S REF. #: (B)(4).