NIM® EMG ELECTRODE
Received Feb 21, 2024 · Event occurred Jan 26, 2024
Report 9612501-2024-00497 · MDR key 18746238
Device
Generic name
Electrode, Needle
Manufacturer
Covidien LlcModel number
8227410Catalog number
8227410
Lot number
0226600753
Product problems
- Inappropriate Audible Prompt/Feedback
- Inappropriate Audible Prompt/Feedback
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H03: PRODUCT ANALYSIS FOUND THAT VISUALLY, THE PRODUCT WAS RETURNED IN A SMALL ZIP LOCK BAG. THE PACKAGING CARTON WAS DAMAGED WHEN RETURNED. THE PACKAGING CARTON CONTAINED ONLY A BLUE ELECTRODE. THERE WAS NO DAMAGE NOTED IN THE CONSTRUCTION OF THE ELECTRODE. THERE WERE LIGHT TRACES OF CONTAMINATION FOUND ON THE NEEDLES. THE CONTINUITY CHECK WAS PERFORMED AND ELECTRICALLY, THE RESISTANCE BETWEEN PAIRED ELECTRODE CABLES SHALL MEASURE LESS THAN 2.0 OHMS AND THE ACTUAL MEASUREMENT FOR A BLUE ELECTRODE BETWEEN POLES WAS RANGING FROM 7OHMS TO 300 OHMS. THERE WAS NO CONSISTENCY RECORDED DURING THE ELECTRODE CHECK. A REVIEW OF THE GLOBAL COMPLAINT DATA SHOWED NO OTHER COMPLAINTS ABOUT THIS LOT NUMBER. IN THE RETURNED CONDITION, THERE WAS AN OUT OF SPECIFICATION CONDITION THAT WAS RELATED TO THE COMPLAINT. H6: PREVIOUS CODES B21, C21 AND D16 ARE NO LONGER APPLICABLE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Additional Manufacturer Narrative
H3: ANALYSIS RESULTS WERE NOT AVAILABLE AS OF THE DATE OF THIS REPORT. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN ANALYSIS IS COMPLETE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
IT WAS REPORTED THAT DURING CANALPLASY SURGERY, ELECTRODE ON CHANNEL ONE WAS GIVING HIGH SIGNALS DURING SETUP REPLACED BLUE ELECTRODE ON CHANNEL 1 AND THIS ISSUE RESOLVED. THERE WAS A DELAY OF LESS THAN ONE HOUR IN SURGERY. NO PATIENT IMPACT.
Additional Manufacturer Narrative
FDC/ANNEX D CODE HAS BEEN UPDATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.