inforMED
InjuryOCZ

STONE CRUSHING FORCEPS, SGL. ACTION JAWS

Received Feb 16, 2024 · Event occurred Nov 29, 2023

Report 2020550-2024-00039 · MDR key 18722589

Device

Generic name

Stone Crushing Forceps, Sgl. Action Jaws

Model number

27074B

Catalog number

27074B

Product problems

  • Difficult to Open or Close
  • Difficult to Open or Close

Patient

58 YR · Male

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

THE REPORTED DEVICE IS PENDING TO RETURN FOR EVALUATION. ONCE THE REPORTED DEVICE IS RETURNED AND EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WOULD BE MADE TO THE FDA. THIS COMPLAINT WILL BE TRACKED AND TRENDED. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE STONE CRUSHING FORCEPS' TIP WOULD NOT CLOSE DURING CYSTOSCOPY AND CYSTOLITHOPAXY PROCEDURES. DURING THE PROCEDURE, THE SURGEON ATTEMPTED TO REMOVE BLADDER STONE BY USING STONE CRUSHING FORCEPS; HOWEVER, THE FORCEPS MALFUNCTIONED WHILE INSIDE THE PATIENT. THE SURGEON COULD NOT REMOVE THE FORCEPS SAFELY FROM INSIDE OF THE PATIENT; THUS, APERINEAL URETHROTOMY IN THE PENILE URETHRA WAS PERFORMED TO REMOVE THE FORCEPS. IT WAS NOTED THAT THERE WAS NO ISSUE WITH THE FORCEPS PRIOR. THE BLADDER STONE DID NOT REQUIRE TO UTILIZE MORE FORCES THAN USUAL TO CRUSH. IT WAS CONFIRMED NOTHING WAS LEFT INSIDE OF THE PATIENT.