BONE HOOK
Received Feb 14, 2024 · Event occurred Jan 10, 2024
Report 1423507-2024-01006 · MDR key 18707509
Device
Generic name
Bone Hook
Manufacturer
Carefusion Inc.Model number
OS4080-002Catalog number
OS4080-002
Lot number
G03
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
UPON RETURN OF THE DEVICE SUBJECT OF THE REPORTED EVENT, THE LOT NUMBER INDICATED THAT THE BONE HOOK WAS APPROXIMATELY 20 YEARS OLD. NORMAL WEAR AND TEAR WERE VISIBLE ON THE DEVICE. DURING EVALUATION OF THE BONE HOOK, THE REPORTED EVENT OF THE BROKEN PIECE WAS CONFIRMED. THE OBSERVED JAGGED EDGE WHERE THE PIECE OF THE BONE HOOKE BROKE OFF MAY BE ATTRIBUTED TO EXCESSIVE FORCE DURING USE. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ABNORMALITIES WERE NOTED. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE A PIECE OF THE BONE HOOK "BROKE OFF" AND FELL INTO THE PATIENT. NO REPORT OF INJURY.
Additional Manufacturer Narrative
THE BONE HOOK SUBJECT OF THE REPORTED EVENT IS BEING SENT BACK FOR EVALUATION. INVESTIGATION OF THIS EVENT IS CURRENTLY IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.