PATCH
Received Oct 15, 2010 · Event occurred Aug 23, 2010
Report 2124215-2010-17039 · MDR key 1870619
Device
Product problems
- High impedance
- Low impedance
- Insufficient Device Problem Information
Patient
88 YR
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND TWO EPICARDIAL LARGE PATCH LEADS WERE EXHIBITING HIGH OUT OF RANGE SHOCKING IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS AND LOW OUT OF RANGE SHOCKING IMPEDANCE MEASUREMENTS LESS THAN 20 OHMS. THE EPICARDIAL PATCHES HAVE BEEN IMPLANTED FOR APPROXIMATELY 19 YEARS. A BOSTON SCIENTIFIC CRM TECHNICAL SERVICES CONSULTANT RECOMMENDED PERFORMING A 1.1 AND 41 JOULE SHOCK TO TEST THE INTEGRITY OF THE SYSTEM AND THE DEVICE'S ABILITY TO DELIVER A SHOCK. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PATIENT PLANNED TO BE MONITORED. THERE WERE NO PLANS FOR ADDITIONAL INTERVENTION AT THIS TIME.
Description of Event or Problem
APPROXIMATELY TWO YEARS LATER THE EPICARDIAL PATCH WAS RETURNED FOR ANALYSIS DUE TO A PATIENT DEATH FROM CONGESTIVE HEART FAILURE.
Additional Manufacturer Narrative
ATTEMPTS TO RETRIEVE ADDITIONAL INFORMATION HAVE BEEN MADE. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE EVENT WILL BE UPDATED.
Description of Event or Problem
--
Additional Manufacturer Narrative
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE RETURNED LEAD SEGMENT WAS PERFORMED.THE LEAD WAS SEVERED 14.2 CM FROM THE TERMINAL PIN, AND ONLY THE PROXIMAL SEGMENT WAS RETURNED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS.