Received Feb 9, 2024 · Event occurred Jan 18, 2024

Report 1718873-2024-00001 · MDR key 18677307

Device

Generic name

Umbilical Catheter

Catalog number

4273505

Lot number

1230384

Product problems

  • Material Split, Cut or Torn
  • Material Split, Cut or Torn

Patient

NA · Unknown

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Description of Event or Problem

ON 1/18/2024, CATHETER BROKE DURING REMOVAL BY PHYSICIAN ASSISTANT. THE REMAINING CATHETER WAS REMOVED WITH SUB XYPHOID SURGICAL RE-ENTRY.

Additional Manufacturer Narrative

UTMD IS REPORTING THIS EVENT BECAUSE THE HOSPITAL STAFF STATED THAT A SUB XYPHOID SURGICAL RE-ENTRY WAS PERFORMED TO REMOVE THE REMAINING CATHETER FROM THE PATIENT. THE CT SURGERY ADVANCED LEAD AT THE USER FACILITY STATED THAT THE PATIENT IS DOING WELL. AFTER INVESTIGATION AT UTMD, THE RETURNED CATHETER HAD BEEN DAMAGED WITH SHARP INSTRUMENT AND THEN TORN. THE RETURNED CATHETER WAS PULL TESTED AT UTMD AND MET SPECIFICATIONS. THE IFU HAS CAUTIONS INDICATING TO TAKE CARE NOT TO DAMAGE THE SOFT SILICONE CATHETER DURING USE. (SEE IFU 57078 REV 022322).