inforMED
InjuryFML

RITTER

Received Oct 6, 2010 · Event occurred Sep 10, 2010

Report 1523530-2010-00009 · MDR key 1864778

Device

Generic name

Chair, Blood Donor

Manufacturer

Midmark Corp.

Model number

104-035

Product problems

  • Scratched Material

Patient

Not reported

  • Laceration(s)

Narrative

Description of Event or Problem

PT SLID OFF THE FOOT STEP AND CUT HER LEG ON THE FOOT STEP FRONT REQUIRING NINE STITCHES. THERE IS A SHARP EDGE TO THE FOOT STEP FRONT. THE PT WAS NOT VERY MOBILE, REQUIRED A CAREGIVER TO HELP HER.

Additional Manufacturer Narrative

THE EXAM TABLE INVOLVED IN THE INCIDENT WAS 30 YEARS OLD. AN INDEPENDENT SERVICE TECH WAS SENT TO EVALUATE THE TABLE. THE TECH REPORTED THAT THE EDGE BANDING ON THE STEP FRONT WAS MISSING. WE BELIEVE THE MISSING BANDING CONTRIBUTED TO THIS INCIDENT.

Remedial action

  • Inspection