inforMED
MalfunctionBTM

MEDLINE

Received Feb 1, 2024 · Event occurred Aug 25, 2023

Report 1314417-2024-00009 · MDR key 18628109

Device

Generic name

Adult Resus W/mask And Bag Reservoir

Model number

CPRM1116

Catalog number

CPRM1116

Lot number

UNKNOWN

Product problems

  • Inflation Problem
  • Inflation Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE MANUAL RESUSCITATOR INABILITY TO INFLATE WOULD DELAY THE RESUSCITATION OF PATIENT.

Description of Event or Problem

NOT INFLATING.

Additional Manufacturer Narrative

THE MANUAL RESUSCITATOR INABILITY TO INFLATE WOULD DELAY THE RESUSCITATION OF PATIENT. THE COMPLAINT OF "NOT INFLATING" REGARDING PART CPRM1116 WAS NOT CONFIRMED. THE ROOT CAUSE CANNOT BE DETERMINED BUT COULD POSSIBLY BE A RESULT OF "MFG ERROR - CONNECTION TO SQUEEZE BAG LEAKS". A RISK ASSESSMENT WAS PERFORMED AND THE ULTIMATE RISK WAS DETERMINED TO BE LOW WHICH DOES NOT REQUIRE THE INITIATION OF A CAPA. THERE HAVE BEEN 0 OTHER COMPLAINTS WERE REGARDING THE SAME PART AND A SIMILAR ISSUE WITHIN THE 24 MONTHS PRECEDING THIS COMPLAINT. NO RESPONSE WAS REQUESTED FROM THE CUSTOMER.

Description of Event or Problem

NOT INFLATING.