BD MULTITEST
Received Jan 29, 2024 · Event occurred Jan 12, 2024
Report 2916837-2024-00004 · MDR key 18596934
Device
Generic name
Counter, Differential Cell
Manufacturer
Becton Dickinson And Company Bd BiosciencesCatalog number
662967
Lot number
69761
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT WHILE USING BD MULTITEST CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANTIBODIES CONTAMINATED WITH SUBSTANCE THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC , ,CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 69761.
Additional Manufacturer Narrative
INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE (NON-SPECIFIC STAINING CREATING UNUSUAL STAINING PATTERNS FOR CD4+CD+8 POPULATIONS) WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: DATA PROVIDED BY CUSTOMER WAS REVIEWED. RETAIN TESTING RESULTS SHOWED THE PRESENCE OF A NON-SPECIFIC STAINING. POTENTIAL CAUSE IS DETERMINED TO BE MANUFACTURING RELATED. ADDITIONAL CONTROLS HAVE BEEN PUT IN PLACE IN THE MANUFACTURING STAGE SPECIFIC TO THE SUPPLIED COMPONENT. CORRECTIVE ACTION WAS INITIATED TO ADDRESS THE ROOT CAUSE AND PREVENTIVE ACTIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
Additional Manufacturer Narrative
G5. PMA/510K: K170974 H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED THAT WHILE USING BD MULTITEST CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANTIBODIES CONTAMINATED WITH SUBSTANCE THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC , ,CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 69761
Remedial action
- Recall