inforMED
MalfunctionFPB

MICRO-VOLUME EXTENSION SET

Received Jan 29, 2024

Report 1416980-2024-00217 · MDR key 18596799

Device

Generic name

Filter, Infusion Line

Model number

NA

Catalog number

2N3350

Lot number

DR23G10024

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H10: THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. FUNCTIONAL PRESSURE TEST AND CLEAR PASSAGE TESTS WERE PERFORMED; NO LEAK AND BLOCKAGE WERE IDENTIFIED. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED FROM THE FILTER. THE SET WAS CONNECTED TO AN UPPER EXTREMITY PEDIATRIC PICC (PERIPHERALLY INSERTED CENTRAL CATHETER) LINE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.