MalfunctionJIL

CLINITEK STATUS+

Received Jan 22, 2024 · Event occurred Dec 27, 2023

Report 3002637618-2024-00003 · MDR key 18552287

Device

Generic name

Clinitek Status+

Catalog number

10379675

Product problems

  • Overheating of Device
  • Overheating of Device

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SIEMENS REQUESTED THAT THE CUSTOMER RETURN THEIR INSTRUMENT AS WELL AS THE POWER CORD AND BATTERIES IN USE AT THE TIME OF THE EVENT FOR INVESTIGATION. SIEMENS HAS SHOWN DUE DILIGENCE IN REQUESTING THE RETURN OF THE CUSTOMER'S INSTRUMENT. HOWEVER, SIEMENS HAS NOT RECEIVED ANY ADDITIONAL UPDATE ON THE RETURN. WITHOUT THE REQUESTED PRODUCT, THE INVESTIGATION COULD NOT PROCEED. THE CAUSE OF THIS EVENT IS UNKNOWN. IT SHOULD BE NOTED THAT IF THE CUSTOMER WAS USING THE STATUS CONNECT PLATFORM, BATTERIES ARE NOT RECOMMENDED. PER THE INSTRUMENT MANUAL: "CAUTION DO NOT USE BATTERIES IN THE ANALYZER, IF YOU ATTACH THE ANALYZER TO A CLINITEK STATUS CONNECTOR. LEAVING THE BATTERIES IN THE BATTERY COMPARTMENT MAY CORRODE THE BATTERIES."

Description of Event or Problem

THE CUSTOMER REPORTED THAT THEIR STATUS+ INSTRUMENT PRODUCED VISIBLE SMOKE AND A BURNT SMELL. THE CUSTOMER STATED THAT THEY WERE USING BOTH BATTERIES AND A POWER CORD AT THE TIME OF THE EVENT. THE CUSTOMER WAS ADVISED TO IMMEDIATELY UNPLUG THE UNIT AND REMOVE IT FROM SERVICE. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.