inforMED
InjuryIKK

TREXO PLUS

Received Jan 17, 2024 · Event occurred Dec 11, 2023

Report MW5150483 · MDR key 18536150

Device

Generic name

System, Isokinetic Testing And Evaluation

Model number

A0001

Catalog number

A001BPR

Patient

Male

  • Bone Fracture(s)
  • Bone Fracture(s)

Narrative

Description of Event or Problem

DURING THE PATIENT'S THERAPY SESSION AT CLINIC, THE CHILD (WC) WALKED IN THEIR TREXO PLUS DEVICE FOR THE SECOND TIME, AS PART OF THEIR REHABILITATIVE ACTIVITIES. UPON RETURNING HOME MOM NOTICED THE FRACTURE WHICH WAS LATER CONFIRMED ON X-RAY. AFTER HEARING OF THE INCIDENT, COMPANY REPRESENTATIVE WENT TO INVESTIGATE THE DEVICE AT THE CLINIC ON (B)(6) 2023 AT 12:50PM. HE FOUND EVIDENCE OF USER ERROR ON THE DEVICE ITSELF AS A RESULT OF CLINIC SUPERVISORS FAILING TO COMPLY WITH THE TREXO PLUS USER MANUAL, AND/OR TRAINING PROVIDED BY THE TREXO CUSTOMER SUCCESS TEAM.