inforMED
MalfunctionNHW

PATCH

Received Oct 5, 2010 · Event occurred Jul 15, 2010

Report 2124215-2010-17165 · MDR key 1852895

Device

Generic name

Implantable Lead

Manufacturer

Cpi - Del Caribe

Model number

0067

Product problems

  • High impedance
  • Failure to Convert Rhythm
  • Material Integrity Problem

Patient

51 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS EPICARDIAL PATCH EXHIBITED >125 OHMS SHOCK IMPEDANCE IN ALL VECTORS. IT WAS DETERMINED THAT THIS PATCH WHICH WAS PLUGGED INTO THE DISTAL PORT, HAD BECAME FRACTURED. THEREFORE, SHOCK THERAPY COULD NOT BE GUARANTEED. THE PATIENT WAS NOTED AS PACEMAKER DEPENDENT, BUT THERE WAS NO EVIDENCE OF IMPACT ON CRITICAL THERAPY.

Additional Manufacturer Narrative

THIS EPICARDIAL PATCH WAS SURGICALLY ABANDONED. IF NEW INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE FURTHER EVALUATED AND UPDATED.