Narrative

Additional Manufacturer Narrative

INVESTIGATION SHOWED THAT THE CASTING FOR THE LEG SECTION CONNECTOR WAS DAMAGED. THIS PART WAS BENT BEYOND ITS SPECIFICATION AND DID NOT ALLOW THE LEG SECTION TO PROPERLY CONNECT TO THE TABLE. THIS DAMAGED CONNECTOR CONTRIBUTED TO THE EVENT.

Description of Event or Problem