Received Sep 16, 2010 · Event occurred Aug 1, 2010

Report 2244861-2010-00001 · MDR key 1841305

Device

Generic name

Pressure Infusor For An Iv Bag

Model number

803FGA

Catalog number

803FGA

Lot number

UNK

Product problems

  • Occlusion Within Device
  • Decrease in Pressure

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

"THE INFORMATION PASSED ON FROM THE DISTRIBUTOR INDICATES THAT THE DEVICE USED IN THE HOSPITAL DID NOT MAINTAIN ITS DESIRED INFLATION PRESSURE FOR THE EXPECTED TIME OF USE. GRADUALLY DECREASING INFLATION PRESSURE LED TO THE OCCLUSION OF A PATIENT'S ARTERIAL LINE AND RESULTED IN THE NEED TO REPLACE THE ARTERIAL LINE. THERE WAS NO ADVERSE EFFECT TO THE PATIENT. THE PRODUCTS GAUGE T-FITTING WAS IDENTIFIED BY VISUAL EXAMINATION AS THE SOURCE OF THE LEAK.

Remedial action

  • Replace