inforMED
MalfunctionIXL

DIAGNOST 56/66/76

Received Sep 9, 2010

Report 3003768251-2010-00023 · MDR key 1839783

Device

Generic name

Ixl (device, Spot Film)

Model number

70641

Catalog number

NA

Product problems

  • Bent
  • Break
  • Detachment Of Device Component

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

THE COLLIMATOR WAS HANGING BY THE DETENT PIN THAT WAS STUCK ON THE LIP OF THE COLLIMATOR BRACKET. THE DETENT MECHANISM WAS BENT AND DAMAGED. THE TOP COVER WAS LYING ON THE COUNTER.

Additional Manufacturer Narrative

THE SCREWS WERE REPLACED AND THE COLLIMATOR WAS REMOUNTED AFTER BENDING THE DETENT ASSEMBLY AND THE OFF FOCAL VANES BACK INTO POSITION. AFTER TROUBLESHOOTING IT WAS FOUND THAT A BREAKER WAS TRIPPED ON THE CONTROL RACK AND THE PIN WAS DAMAGED ON THE MAIN CONNECTOR FOR THE COLLIMATOR, WHICH WAS EVENTUALLY REPLACED.

Remedial action

  • Repair